Feature Options Default Setting Limits, Range of Values
AUC Limits Always on Low Limit
70 (mg/dL)
3.9 (mmol/L)
High Limit
180 (mg/dL)
10.0 (mmol/L)
Low Limit range:
a
Lowest
40 (mg/dL)
2.2 (mmol/L)
Highest
400 (mg/dL)
22.2 (mmol/L)
High Limit range:
b
Lowest
40 (mg/dL)
2.2 (mmol/L)
Highest
400 (mg/dL)
22.2 (mmol/L)
Block
•On
•Off
Off Possible values:
On
Off
Blood Glucose
Units (BG Units)
•mg/dL
•mmol/L
mg/dL Possible values:
mg/dL
mmol/L
Cal Reminder
•On
•Off
On 1 hour Range:
Shortest
0:05 (5 minutes)
Longest
6:00 (6 hours)
a. The AUC Low Limit can be equal to, but not higher than the AUC High Limit.
b. The AUC High Limit can be equal to, but not lower than the AUC Low Limit.
System specifications114