6 CGMS
®
iPro
TM
Continuous Glucose Recorder User Guide
CGMS iPro Magnetic Wand interference
Do not store the CGMS iPro Magnetic Wand or any other magnet within 1.5 inches (3.8
cm) of the CGMS iPro Digital Recorder. Premature battery discharge may occur.
Do not place the CGMS iPro Magnetic Wand or any other magnet within 1.5 inches (3.8
cm) of the CGMS iPro Digital Recorder except while performing patient set-up and data
download procedures. Premature battery discharge may occur.
CGMS iPro Magnetic Wand interference with insulin pumps
Do not place the CGMS iPro Magnetic Wand directly on any insulin pump. Incorrect
pump operation may occur.
Do not insert the glucose sensor with attached CGMS iPro Digital Recorder closer than
two to three inches (5.08 - 7.62 cm) from the insulin pump. The use of the CGMS iPro
Magnetic Wand may inadvertently cause incorrect operation of the pump.
CGMS iPro Magnetic Wand interference with PC hard drives
Do not place the CGMS iPro Magnetic Wand on your PC or external hard drives, as the
magnetic field may impact the integrity of the data files stored there.
Other electromagnetic interference
The CGMS iPro Digital Recorder is designed to withstand common electromagnetic
interference, including airport security systems.
Precautions
If performing multiple CGMS iPro Digital Recorder studies on the same patient, establish
a rotation schedule for choosing new glucose sensor sites. Avoid sites that are
constrained by clothing, have scar tissue, or are subject to rigorous movement during
exercise.
If any blood gets inside the digital recorder connector, the digital recorder must be
discarded.
If you do not plan to use the digital recorder for at least 30 days, connect it to the
charger for storage. The charger will not keep the digital recorder charged, but it will
slow the digital recorder’s battery discharge. Remove the digital recorder from the
charger at least once every 60 days, wait at least one minute, then reconnect the
digital recorder to the charger. To use the digital recorder after it has been stored,
disconnect it from the charger, wait at least one minute, then reconnect it again to
start the charging process. It may take up to eight hours to fully charge a digital
recorder that has been stored for more than 30 days.