Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The iPro2 CGM system is intended for use in the electromagnetic environment specified below. The customer
or user of the iPro2 CGM system should assure that it is used in such an environment.
Immunity Test
IEC 60601
Level
Compliance Level
Electromagnetic Environment Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the iPro2 CGM system, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
d=1.2 P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
6.0 GHz
3 V/m
d=1.2 P
80 MHz to 800 MHz
d=2.3 P
800 MHz to 6.0 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption,
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the iPro2 CGM system is
used exceeds the application RF compliance level above, the iPro2 CGM system should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the iPro2 CGM system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
iPro2 CGM User Guide Specifications and notices 89