The arrow direction shows if your SG measurements are rising or falling. The number of arrows (one or two)
shows how rapidly your SG is changing.
One up arrow
shows that your SG has been rising at a rate of
1 to 2 mg/dL (0.05 to 0.11 mmol/L) per minute.
240
3:16P
3 Hour
One down arrow shows that your SG has been falling at a rate
of 1 to 2 mg/dL (0.05 to 0.11 mmol/L) per minute.
3:11P
131
3 Hour
Two up arrows show that your SG has been rising at a rate
of 2.0 mg/dL (0.11mmol/L) or more per minute.
3:11P
212
3 Hour
Two down arrows show that your SG has been falling at a
rate of 2.0 mg/dL (0.11 mmol/L) or more per minute.
3:11P
103
3 Hour
Using your sensor
135