EMPTY RESERVOIR
There is no insulin in the reservoir. Change the reservoir immediately.
FAILED BATT TEST
The pump tests the voltage of each battery installed. This test ensures a battery with low voltage is not used.
If the battery does not have enough voltage, this alarm will occur. The pump will not function and the battery
must be replaced. (Always make sure that you install a NEW battery into the pump.)
MAX DELIVERY
This alarm alerts you when you have taken more insulin than expected based on Maximum bolus and Maximum
basal rates.
MOTOR ERROR
Insulin delivery has stopped. This alarm occurs if your pump detects a motor error.
NO DELIVERY
Insulin delivery has stopped. This alarm occurs if your pump detects a blockage.
NO RESERVOIR
The reservoir is not inserted correctly or no reservoir has been inserted.
OFF NO POWER
The battery is dead. Replace battery immediately. Follow the directions on the screen. Check to make sure
that the time is correct on the screen. Reset the time if necessary.
Troubleshooting and alarms
169
Chapter 13