Medtronic Paradigm Link Guía de usuario Pagina 201

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 262
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 200
Delivery accuracy
Delivery accuracy: +/-5 percent.
Displacement (piston) accuracy: +/-2 percent.
The delivery accuracy test was done using the Sof-Set
®
Ultimate infusion set (MMT–315).
-50
-40
-30
-20
-10
0
10
20
30
40
50
0 15 30 45 60 75 90 105 120 135 150 165 180 195 210
Maximum Error (percent)
Minimum Error (percent)
Ovreall Flow Error (percent)
Time Interval (Minutes)
Percent Error
Delivery Volume Accuracy Curve (-0.37%) Overall Error
(MMT-522 Model) Qualification Testing (0.01 mL/hr 09/22/04 S/N: 010112-C061); 23°C
Pump specifications
182
Vista de pagina 200
1 2 ... 196 197 198 199 200 201 202 203 204 205 206 ... 261 262

Comentarios a estos manuales

Sin comentarios