Selecting the BG units
Y
ou must select the blood glucose units (BG Units) for the system (either mmol/L or mg/dL). All BG
measurements will show in the BG unit type you select.
If you have turned on your Bolus Wizard feature, BG units will not be available for sensor setup. Please go
to the Setting the BG units section of the Using the Bolus Wizard feature chapter for details.
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2 Select BG Units, then press ACT. The
BLOOD GLUCOSE UNITS screen shows mg/dL and mmol/L.
3 Select your BG units.
4 Press ACT. The EDIT SETTINGS screen shows the BG Units type you selected.
You are now ready to set up your High Glucose limit.
High Glucose Alert
Y
our High Glucose alert must be turned on if you want your pump to alert when the sensor glucose
measurement reaches or goes above your set High Glucose limit.
To turn on your High Glucose alert, and to set the High Glucose limit:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2 Select High Glucose, then press ACT.
3 Select On, then press ACT. The
SET HIGH GLUCOSE screen flashes the default High Glucose limit of
200 mg/dL (11.2 mmol/L).
4 Select your High Glucose limit between 110–400 mg/dL (2.8–22.2 mmol/L). Your High Glucose limit
must be at least 10 mg/dL (0.6 mmol/L) above your Low Glucose limit.
5 Press ACT. The EDIT SETTINGS screen shows the High Glucose value you selected. You are now ready
to set up your High Snooze.
Sensor features112