Stopping your pump
Suspend stops all insulin delivery including the current basal and any bolus or prime deliveries that are in
progress. While suspended, your pump will not deliver insulin until you resume your pump. When you
resume your pump, the basal delivery will continue.
The pump will beep or vibrate about every 15 minutes on the hour to remind you that it is not delivering
insulin. Example: Y
ou suspend your pump at 11:20 a.m. The pump will beep or vibrate at 11:30 a.m.,
11:45 a.m., 12:00 noon, and so on until you resume your pump (basal resumes).
When suspended, your pump is in Attention mode with a solid circle showing on the pump. When in
Suspend, you can only resume your basal or view the STATUS screen. No other functions are available.
To suspend the pump:
1 Select Suspend from the
MAIN MENU, and press ACT.
Main > Suspend
2 SUSPEND will flash on your screen. Press ACT to stop your pump.
3 The screen will show that the pump is suspended and the time that it stopped. After less than one
minute, the pump will return to the HOME screen with a solid circle.
You can verify on your pump STATUS screen that your pump is suspended.
Resume pump delivery
When the pump is suspended, it defaults to the HOME screen with a solid circle.
T
o resume your pump and basal delivery, do these steps:
1 From any screen, press ACT until the
RESUME screen appears.
Press ACT again.
2 Your pump will beep once, then the HOME screen will appear without the solid circle.
NOTE: A bolus or fixed prime that was stopped by suspend will not restart when you resume your
pump. Y
ou must reprogram and activate it to finish delivery.
Basic programming 55