Sensor features default settings
Feature Options Default Setting Limits, Range of Values
Sensor
On
Off
Off None
High Glucose
Alert
On
Off
Off Low–22.2 mmol/L (Low–400
mg/dL)
a
Low Glucose
Alert
On
Off
Off 2.2 mmol/L–High
(40mg/dL–High)
b
Alarm Snooze (Always on - cannot be turned
off)
On 30 minutes Range: 0:05 - 1:00 (5
minutes - 1 hour)
Cal Reminder
On
Off
On 1 hour Range: 0:05 - 4:00 (5
minutes - 4 hours)
Blood Glucose
Units (BG Units)
mg/dL
mmol/L
mg/dL None
Missed Data (Always on - cannot be turned
off)
30 minutes Range: 0:05 - 0:40 (5
minutes - 40 minutes)
High Snooze 1:00 (1 hour) Range: 0:05 - 3:00 (5
minutes - 3 hours)
Low Snooze 0:20 (20 minutes) Range: 0:05 - 1:00 (5
minutes - 1 hour)
a. Your High Glucose limit must be at least 0.6 mmol/L (10 mg/dL) above your Low Glucose limit.
b. Your Low Glucose limit must be at least 0.6 mmol/L (10 mg/dL) below your High Glucose limit. For example, if your High
Glucose limit is 11.1 mmol/L (200 mg/dL), then your Low Glucose limit cannot be higher than 10.5 mmol/L (190 mg/dL).
Pump specifications196