Medtronic Paradigm 522 Guía de usuario Pagina 229

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 262
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 228
The Clarke Error Grid was used to assess the clinical relevance of the differences between the Guardian RT
readings and the comparative YSI measurements. The Clarke Error Grid divides a correlation plot into 5
zones. See following table.
R
esults in zones A and B are considered clinically acceptable, while results in zones C, D, and E are
potentially dangerous and, therefore, clinically significant errors. The Clarke Error Grid zones are labeled
on the correlation plot.
Zone Description
A Clinically accurate, would lead to correct treatment decisions
B Would lead to benign decisions or no treatment
C Would lead to over-correction of normal glucose levels
D Would lead to failure to detect and treat high or low glucose levels
E Would lead to erroneous treatment decisions
Appendix A210
Vista de pagina 228
1 2 ... 224 225 226 227 228 229 230 231 232 233 234 ... 261 262

Comentarios a estos manuales

Sin comentarios