Item When What
Standard basal delivery data (always appears) current basal rate (basal 1, basal 2,
etc.)
Status of pump (such as Rewind, Suspended, Low Reservoir, Set Time, and so on)
Time (always appears)
*If all of the Reminders and the Alarm Clock are enabled, only the reminder that is nearest to alarming
will appear in the ST
ATUS screen.
Bolus Wizard feature specifications
There are three different formulas the Bolus W
izard feature uses to estimate a bolus, depending on your
current BG. The following formulas apply only when the carb units are in grams.
1 If your current BG is greater than your High BG Target, the Bolus Wizard feature subtracts active
insulin from the BG correction estimate, then adds this to the food estimate to get the total bolus
estimate. However, if the result of subtracting active insulin from BG correction estimate is a negative
number (less than zero), the total bolus estimate is based only on the food estimate.