CAL ERROR
Reason: An error occurred when entering a new meter BG measurement to calibrate the system. Some
possible causes are:
• An incorrect blood glucose number was entered from the meter into the pump.
• The entered blood glucose measurement was not current.
• Your blood glucose is rising or falling rapidly.
• The sensor needs more time to stabilize after being inserted.
• The sensor is no longer reading the sensor glucose correctly.
Action:Follow the guidelines in the Calibrating the sensor section in the Sensor features chapter to enter
a new meter BG for calibration. If you receive a Cal Error on your second calibration, a CHANGE SENSOR
alert will occur. Call our 24 Hour HelpLine if you have questions.
METER BG NOW
Reason: A meter BG measurement is needed immediately to calibrate your sensor so that you can
continue receiving sensor glucose readings.
Action: Take and enter a meter blood glucose measurement. Follow the guidelines in the Calibrating the
sensor section in the Sensor features chapter.
METER BG BY
Reason: A meter BG measurement must be entered by the time that is shown to calibrate your sensor. The
METER BG BY alert is also known as the CAL REMINDER alert.
Action: Enter a blood glucose measurement to avoid a METER BG NOW alert.
LOW XX MG/DL (XX = SG measurement)
Reason: The sensor glucose number is equal to or lower than your low glucose limit. If you do not set a
Low Glucose Limit and do not set the glucose alerts ON, then you will not get a Low sensor glucose alert.
The pump plays four consecutive tones, in falling pitch, if an audible beep has been selected as the alert
type.
Troubleshooting, alerts and alarms162