Bolus Wizard feature default settings
Item Default
Setting
Limits Increments Warning Limits
carb units: grams — — —
ins to carb (or
exch) ratio:
15 grams/u
or
1 unit/exch
1–200 g/u or 0.075–
15.0 u/exch
0.1 g/u for 0–9.9 g/u;
1 g/u for ratios of 10.0 g/u
or more
(or 0.001 u/exch for 0.075–
0.099 u/exch; 0.01 u/exch
for 0.10–9.99 u/exch;
0.1 u/exch for 10.0–
15.0 u/exch)
5–50 g/u or 0.3–
3.0 u/exch
(insulin)
sensitivity:
50 mg/dL/u
or
2.8 mmol/L/
u
10–400 mg/dL/u or
0.5–22.2 mmol/L/u
1 mg/dL/u or
0.1 mmol/L/u
20–100 mg/dL/u
or 1.1–
5.6 mmol/L/u
BG Target: 100–
100 mg/dL or
5.6–
5.6 mmol/L
60–250 mg/dL or
3.3–13.9 mmol/L
1 mg/dL or 0.1 mmol/L 90–140 mg/dL or
5.0–7.8 mmol/L
Active Ins
Time:
6 hours 2–8 hours 1 hour —
Sensor features default settings
Feature Options Default Setting Limits, Range of Values
Sensor
•On
•Off
Off —
Glucose Alerts
•On
•Off
Off Set the limits using the
Glucose Limits feature.
Pump specifications194