Meter
The Paradigm pump can be programmed to receive fingerstick blood glucose values transmitted from
Bayer's CONTOUR
®
NEXT LINK Wireless Meter, Bayer's CONTOUR
®
LINK Wireless Meter, a OneTouch
®
UltraLink
®
meter, or a BD Paradigm Link
®
glucose meter for use in the Bolus Wizard calculator and for use
as calibration values for the glucose sensor. Glucose values from other glucose meters must be manually
entered into the pump.
Contraindications
Pump therapy is not recommended for people who are unwilling or unable to perform a minimum of four
blood glucose tests per day and to maintain contact with their healthcare professional. Successful insulin
pump therapy requires sufficient vision or hearing to allow recognition of the pump signals and alarms.
Do not expose your insulin pump to MRI equipment or other devices that generate very strong magnetic
fields. The magnetic fields in the immediate vicinity of these devices can damage the part of the pump’s
motor that regulates insulin delivery, possibly resulting in over-delivery and severe hypoglycemia.
Your pump must be removed and kept outside the room during magnetic resonance imaging (MRI)
procedures.
If your pump is inadvertently exposed to a strong magnetic field, discontinue use and contact the 24 Hour
HelpLine for further assistance.
Warnings
Pump
The pump is not suitable for use in the presence of a flammable anaesthetic mixture with air, oxygen or
nitrous oxide.
The drive support cap should appear slightly indented at all times
(as shown in the image). If you notice that the cap is loose, or
sticking out, discontinue use of the pump and contact our
24 Hour HelpLine at 1 800 646 4633 to arrange for your pump to
be replaced. Do not press on the drive support cap if it is
sticking out. Pushing on the drive support cap may result in
unintended delivery of insulin, which can cause hypoglycemia.
drive support cap