Medtronic Revel Guía de usuario Pagina 64

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 265
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 63
12:00 AM
Basal Rate 1
start time
3:00 AM
Basal Rate 2
start time
7:00 AM
Basal Rate 3
start time
9:00 PM
Basal Rate 4
start time
24 hours
rate 3
rate 4 rate 1
rate 2
The start and stop times of basal rates for this example are:
12:00 a.m. (00:00) to 2:59 a.m. (2:59) for basal rate 1
3:00 a.m. (3:00) to 6:59 a.m. (6:59) for basal rate 2
7:00 a.m. (7:00) to 8:59 p.m. (20:59) for basal rate 3
9:00 p.m. (21:00) to 11:59 p.m. (23:59) for basal rate 4
If you need one basal insulin that starts before midnight and stops after midnight, you need to set up two
basal rates. These basal rates have the same rate but different start and stop times. For example, in this
graph a basal rate is needed from 9:00 p.m to 2:59 a.m. (21:00 to 2:59) at one rate. Basal rate 1 always
starts at midnight and should be set to end at 2:59 a.m. (2:59). After setting basal rates 2 and 3, rate 4 is
set from 9:00 p.m. (21:00) to 11:59 p.m. (23:59) at the same rate as basal rate 1. This gives continuous
basal insulin from 9:00 p.m to 2:59 a.m. (21:00 to 2:59) at one rate.
You cannot set a start time for one basal rate to overlap the next basal rate. The addition of a new basal
rate will erase any basal rates that follow.
For best results, setting or changing your basal rate(s) should be discussed with your healthcare
professional.