The percent of Guardian RT readings in the previous graph are presented in the following table according
to the percentage of points falling within each zone (A-E). Results are further broken down (stratified)
according to the range of glucose concentrations.
Glucose
Range
(mg/dl)
Number and
(%) of Data
Points
Evaluated
A + B A B C D E
40-80 356 (9) 271 (76.1) 214 (60.1) 57 (16.0) 2 (0.6) 80 (22.5) 3 (0.8)
81-120 769 (20) 768 (99.9) 463 (60.2) 305 (39.7) 1 (0.1)
N/A
a
N/A
121-240 2362 (60) 2352 (99.6) 1476 (62.5) 876 (37.1) 4 (0.2) N/A 6 (0.2)
>240 454 (11) 394 (86.8) 277 (61.0) 117 (25.8) N/A 59 (13.0) 1 (0.2)
Overall 3941 (100) 3785 (96.0) 2430 (61.7) 1355 (34.4) 7 (0.2) 139 (3.5) 10 (0.2)
a. N/A means that the Clarke Error Grid does not consider the possibility of these zones in that concentration range.
Appendix A212