
WARNING: Do not insert the reservoir in the pump if you did not rewind. Doing so could result
in inaccurate insulin delivery.
Never insert the reservoir into the pump while the tubing is connected to your body. Doing so
could result in an accidental infusion of insulin.
2 Insert the reservoir into the top of the pump case.
3
Turn the tubing connector approximately 1/2 -turn clockwise until the connector is seated. The tubing
connector should be aligned horizontally with the pump case as shown here.
tubing connector
2
3
4 Attach the activity guard, if desired.
5 If the pump has returned to the HOME screen, press ACT to show the REWIND COMPLETE screen. Press
ACT again to go to the FILL TUBING screen.
6 You must now fill the infusion set tubing as described in the next section.
Filling the tubing
You need to fill the infusion set tubing with insulin before you insert the set into the body.
WARNING: Make sure the infusion set is disconnected from your body before you fill the infusion set
tubing. Never insert the reservoir into the pump while the tubing is connected to your body. Doing so
could result in an accidental infusion of insulin.
1 After you rewind your pump, the FILL TUBING screen will appear.
a. If the infusion set is NOT disconnected from your body, press ACT to select No on the screen.
After you disconnect the infusion set from your body, press ACT on the next screen to continue.
b. If the infusion set is disconnected from your body, select Yes on the FILL TUBING screen. Press
ACT to continue to the next FILL TUBING screen.
2 Press and hold ACT. The pump will beep six times indicating that the pump is positioning the
reservoir.
Starting on insulin56
Comentarios a estos manuales