Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 523/723/523K/723K pump is intended for use in the electromagnetic environment
specified below. The customer or user of the Paradigm 523/723/523K/723K pump system should
make sure that it is used in such an environment.
Immunity
Test IEC 60601
Level
Compliance
Level Electromagnetic Environment Guidance
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
6.0 GHz
10 V/m
d = 0.35
P
80 MHz to 800 MHz
d = 0.70
P
800 MHz to 6.0 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Pump specifications 201